Show simple item record

dc.contributor.author Mikkelsen, KB
dc.contributor.author Tabar, YR
dc.contributor.author Kappel, SL
dc.contributor.author Christensen, CB
dc.contributor.author Toft, HO
dc.contributor.author Hemmsen, MC
dc.contributor.author Rank, ML
dc.contributor.author Otto, M
dc.contributor.author Kidmose, P
dc.date.accessioned 2023-04-20T06:37:18Z
dc.date.available 2023-04-20T06:37:18Z
dc.date.issued 2019
dc.identifier.citation Mikkelsen, K. B., Tabar, Y. R., Kappel, S. L., Christensen, C. B., Toft, H. O., Hemmsen, M. C., Rank, M. L., Otto, M., & Kidmose, P. (2019). Accurate whole-night sleep monitoring with dry-contact ear-EEG. Scientific Reports, 9(1), 16824. https://doi.org/10.1038/s41598-019-53115-3 en_US
dc.identifier.issn 2045-2322(Online) en_US
dc.identifier.uri http://dl.lib.uom.lk/handle/123/20895
dc.description.abstract Sleep is a key phenomenon to both understanding, diagnosing and treatment of many illnesses, as well as for studying health and well being in general. Today, the only widely accepted method for clinically monitoring sleep is the polysomnography (PSG), which is, however, both expensive to perform and influences the sleep. This has led to investigations into light weight electroencephalography (EEG) alternatives. However, there has been a substantial performance gap between proposed alternatives and PSG. Here we show results from an extensive study of 80 full night recordings of healthy participants wearing both PSG equipment and ear-EEG. We obtain automatic sleep scoring with an accuracy close to that achieved by manual scoring of scalp EEG (the current gold standard), using only ear-EEG as input, attaining an average Cohen's kappa of 0.73. In addition, this high performance is present for all 20 subjects. Finally, 19/20 subjects found that the ear-EEG had little to no negative effect on their sleep, and subjects were generally able to apply the equipment without supervision. This finding marks a turning point on the road to clinical long term sleep monitoring: the question should no longer be whether ear-EEG could ever be used for clinical home sleep monitoring, but rather when it will be. en_US
dc.language.iso en en_US
dc.publisher Nature Portfolio en_US
dc.title Accurate whole-night sleep monitoring with dry-contact ear-EEG en_US
dc.type Article-Full-text en_US
dc.identifier.year 2019 en_US
dc.identifier.journal Sci Rep en_US
dc.identifier.issue 1 en_US
dc.identifier.volume 9 en_US
dc.identifier.database Pub Med en_US
dc.identifier.pgnos 16824 en_US
dc.identifier.doi 10.1038/s41598-019-53115-3 en_US